How Do You Use a VNA?
September 14, 2023Introduction
To answer this question, we should ask a different question first: What problem am I trying to solve and what information do I need to solve it?
Your task may be to evaluate the performance of an RF filter or an RF amplifier, or maybe you need to determine the health of a transmission line which is buried underground or feeding an antenna at the top of a tower. Each of these tasks may be accomplished with a Vector Network Analyzer (VNA), but the settings will be different for each. Let’s look at some of these applications.
Measuring a Filter
Any measurement for a project should be documented properly. In your Engineering logbook, record the following:
- Purpose of the test
- Date and location of the test
- Test environment and conditions
- Model number and type of test equipment used to make the measurement
- Setup diagram or picture
- Notes or comments about the measurement
- Any observed anomalies
- Chart of results or location of data
- Comparison to expected or simulated results
Next, power on the VNA and allow it to warm up for 30 minutes. Connect it to the computer via USB and connect test cables to Ports 1 and 2 of the VNA, tightening with a torque wrench. Run the User Interface (UI) software. It will default to a full frequency sweep and display S11 log magnitude. To verify the test cables, attach a mechanical calibration load to the end of the Port 1 cable using a torque wrench and then gently move it about. The S11 measurement should show a reasonably low value, such as -20 dB as in Figure 1, and the trace should not jump around with cable movement.
Move the calibration load to the end of the Port 2 cable and change the measurement to S22 by clicking on S11 at the top left of the screen and choosing it from the pull-down menu. Move the cable about gently and ensure there is no significant change in the S22 measurement.
Figure 1- Measurement of a Calibration Load for Test Cable Verification
If there is any significant change in the S11 or S22 measurement, the offending cable should be discarded and replaced. In a production test environment, this verification should be performed at the beginning of each shift.
Next, attach the test cables to the filter to be measured. Using the menu on the right-hand side of the UI, set the number of traces to 2— Display/Num of Traces [2]. Trace 2 will automatically default to S21, which is desired. Observe the measurement of the filter on the screen and set the start and stop frequencies to cover the range of interest. Increase the number of points as desired for more resolution. If the noise floor needs to be lower to measure a deep stopband, decrease the IF bandwidth and increase the stimulus output level to maximum. Change the scale of the S21 measurement to see a stopband less than 100 dB at the same time as the insertion loss if necessary.
Detach the filter from the cables from the Device Under Test (DUT) and perform calibration. From the calibration menu, choose the desired Cal Kit. Perform calibration, choosing full 2-port SOLT calibration for best accuracy. If an Automatic Calibration Module (ACM) is available, use that instead, as it will produce a superior calibration compared to a mechanical kit and will ensure the best accuracy. See [1] for more details on calibration.
When calibration is complete, reattach the filter and make the desired measurements.
Figure 2- Filter Measurement
The Bandwidth Search feature from the Marker Math menu was turned on for the S21 measurement, so all important filter statistics are visible. Markers 2 and 3 show stopband performance better than 110 dB, and the return loss is also better than 20 dB from 3.87 to 4.10 GHz on the S11 trace. Record the results of the measurement in your logbook.
Measuring an Amplifier
Set the start and stop frequencies, as well as the number of points appropriate for the measurement and perform full 2-port calibration. Next, reduce the VNA stimulus power such that the output level from the amplifier is around 0 dBm. Allowing the level to exceed this may result in receiver compression and the measurement results will be inaccurate. The VNA calibration will still be accurate after the power change. Set the number of traces to 4, measuring S11, S12, S21, and S22, and allocate them in a quad configuration as in Figure 3. Add markers as needed.
Figure 3- Amplifier Measurement
The four measurements –S11, S12, S21, and S22— depict input return loss, reverse isolation, gain, and output match: all important characteristics of an amplifier.
You can perform a power sweep measurement at a single frequency. This is convenient to measure 1 dB compression, P1dB. If the compression level is higher than 0 dBm, pad the output of the amplifier. The procedure for this and other amplifier measurements is outlined in more detail in [2].
Measuring an Antenna and Feedline
For preventative maintenance, measure the return loss of a transmit or receive antenna mounted on a tower. The antenna can be 100 feet or more above the ground, so it is more desirable to evaluate from the safety of the ground. First, measure the return loss of the tower mounted antenna by measuring S11 at the beginning of the cable feedline. Connect VNA Port-1 to the feedline directly or through a good quality test cable connected to it. Calibrating the VNA is not strictly necessary, as it has been calibrated to the output connector by the factory and, as long as the test cable has good return loss, it needn’t be calibrated out.
Antenna return loss should be measured in the operating bandwidth and compared to allowed operating parameters for the site. Figure 4 shows the measurement of return loss and VSWR of a UHF Yagi antenna. A limit line was set on the VSWR chart to check for a worst case of 1.7:1. Return loss and VSWR are simply different ways of portraying the same reflection information, but VSWR—Voltage Standing Wave Ratio— is commonly used for antenna measurement.
Figure 4- UHF Antenna Verification
Note that a lossy feedline will artificially improve the VSWR measurement, since some of the reflection is dissipated as heat and not captured by the VNA.
To analyze the feedline itself, set the VNA measurement range as broad as possible with many measurement points—around eight to ten thousand. As an example, for an S5180B 18 GHz VNA, set the start frequency to 2 MHz and the stop to 18 GHz. Set the number of points to 9,000. To be able to measure return loss or characteristic impedance of the feedline vs distance, use the Lowpass Step Time Domain analysis mode. Lowpass mode requires that each measurement point be harmonically relate — in this case, 2, 4, 6, 8…. 18,000 MHz. The start frequency was chosen to be the stop frequency divided by the number of points. A 1 MHz start frequency and 18,000 points would also work, but the sweep would be much slower.
From the main menu, navigate to Analysis>Time Domain. Turn Time Domain on by clicking the button. Set the units to “Feet”. Set the Reflection Type to “One Way”. Click Cable Correction and enter the velocity factor of the cable or select your cable from the Select Cable menu. Make sure that Cable Correction is ON. Return to the previous menu and set the Start and Stop to sensible distances, setting the “Type” to Lowpass Step. Note that the Lowpass Step and Impulse modes require the broadband setup described above. If Time Domain measurements must be made through a bandpass filter, use the bandpass mode and set start and stop frequencies within the passband. Time Domain resolution will be reduced, and it will not be possible to measure return loss or characteristic impedance vs distance, only impulses which represent changes in either measurement.
With the measurement set to S11 Log Mag, you should now see a plot of return loss vs distance along the cable. The return loss shown is the differential reflection of each tiny time slice of cable. The time resolution for an 18 GHz measurement is 1/[2*(Fstop-Fstart)] or 27.8 pS per point. For 9,000 points, that is 250 nS maximum range. If the velocity factor of the cable is 0.9, that equates to a maximum measurement distance of 221.5 feet. If more distance is needed, change the start frequency to 1 MHz and the number of points to 18,000 and the maximum distance will double to 443 feet.
In Figure 5, a 3.75-foot cable is analyzed over its length. The return loss bump due to the mismatch of the cable connector is clearly visible at the beginning. The cable is open on the other end, so the return loss goes to zero as expected. If this were a long cable run for a tower mounted antenna, it would be possible to see return loss degradation near the antenna due to water ingress if it exists. This is a common problem for outdoor antennas and is easily detected in the Time Domain mode.
Figure 5- Time Domain, Return Loss vs Distance in Feet
From the main menu, navigate to Analysis>Conversion and set Conversion to ON. Then, change Log Mat to Lin Mag from the trace type in the upper left corner of the screen. Click on the 0.100 U right after Lin Mag and set it to 5 for 5 ohms per division. Click on the 0.000 U after the slash and change it to 50 to set the reference to 50 ohms. You should now see the characteristic impedance of the cable vs distance. Water ingress would drastically increase the dielectric constant and result in a much lower characteristic impedance near the antenna.
Figure 6- Characteristic Impedance vs Distance in Feet
The Time Domain setup as described here is somewhat complicated, but there are a number of resources on the Copper Mountain Technologies (CMT) website to help explain it. See references [3], [4], and [5].
Conclusion
We examined three use cases for Vector Network Analyzer measurement, which should help a new user understand the basics of VNA measurement. The Time Domain measurement, a feature which comes standard on all CMT VNAs except for the M series, is not exactly a basic measurement, but it was included here because of its importance for feedline evaluation.
References
[1] Brian Walker, “VNA Calibration Theory Introduction”, https://coppermountaintech.com/an-introduction-to-vna-calibration-theory/
[2] Brian Walker, “Amplifier Measurements”, https://coppermountaintech.com/amplifier-measurements/
[3] Video, “Time Domain Analysis with a VNA”, https://coppermountaintech.com/video/time-domain-analysis-with-a-vna/
[4] Video, “Time Domain Reflectometry Measurements with a Vector Network Analyzer”, https://coppermountaintech.com/video/time-domain-reflectometry-measurements-with-a-vector-network-analyzer/
[5] “Time Domain Analysis with a CMT VNA”, https://coppermountaintech.com/time-domain-analysis-with-copper-mountain-technologies-vector-network-analyzers/
Related Post

Can I Use a VNA as a Signal Generator?
September 12, 2023
Yes! We often are asked about using a vector network analyzer for another purpose, as test equipment budgets may need to stretch. But the answer to this question is yes, you can use a VNA as an additional signal source for testing. Below are some basics of how to do it but there is a more robust application note available here to review for more details.You can set the sweep range to “Zero Span” and set the frequency to the desired value and the VNA will output a constant frequency tone. However, there are some caveats. The Compact Series from CMT uses fractional-N phase locked loops to generate signals with high resolution, thus the output frequency may be off by a fraction of a hertz. Therefore, if the vector network analyzer is used as a frequency source, another signal generator is used to generate that same frequency, and the two instruments are provided with the same 10 MHz time-base reference on the rear panel, the VNA signal may not be frequency locked to the generator. However, this would be an unusual requirement.It is generally not necessary for a VNA to hit frequencies precisely. It is only important that the measured internal IF frequencies fall within the IF bandwidth of the DSP filter implemented in the FPGA.The Cobalt Series VNAs, either 9 or 20 GHz use Direct Digital Synthesis (DDS) to generate the fine frequency steps so in that case the output frequencies will be exact.How To Use a VNA as a Signal GeneratorThe simplest way to configure the analyzer as a signal generator is simply to issue a Preset (System > Preset using the menus along the right side of the user interface), then set the Span to 0 Hz, and finally set the Center frequency and output power to the desired signal source settings.To change the output port of the signal from Port 1 to Port 2, simply change the measured S-parameter from S11 to S22. If you need not retain your calibration and display settings, that’s all there is to it! If you do need to keep such settings, however, the sections that follow walk through the steps necessary to generate a CW starting from typical instrument S-parameter sweep settings without a Preset.When using the VNA as a signal source, it’s important to keep in mind certain limitations which might affect its suitability depending on your application:The generator of the VNA will exhibit harmonics of the fundamental frequency as high as its specified harmonic distortion. You can find the specified harmonic distortion of each CMT VNA in its corresponding datasheet; typically, those are approximately -25 dBc.There will also be non-harmonic distortion, typically lower in power than the harmonic distortion.The generated signal’s output power will be subject to the output power accuracy specification of the instrument; +/- 1.0 dB is a typical specification. For more precise output powers, a power meter should be used to confirm or adjust the VNA output power.Some output power levels for CMT’s 1-Port VNAs are High/Low, specified as “Typical Only” and can’t be adjusted. Output power will need to be set with external attenuators or amplifiers according to the application at hand.If sweeping is halted by setting the trigger mode to HOLD, don’t forget that changes made to the output power or frequency will not take effect until a sweep is initiated! You can click on “Single” to apply the new settings and automatically return to HOLD mode after the sweep completes.For those engineers curious about additional uses for a vector network analyzer, CMT has additional resources to explain the options and pros and cons for each. For example, spectrum analyzer vs network analyzer; what are the differences and can a VNA be used as a spectrum analyzer is expanded in this application note.

Why Warm Up a VNA Before Use
May 25, 2023
Engineers regularly face difficult deadlines. In a rush to complete a series of measurements, it is tempting to take a piece of test gear off the shelf, plug it in, and begin the process. Resist the urge! Here, we’ll look at how a typical Vector Network Analyzer (VNA) measurement can change from initial power-up to thermal equilibrium. Note, we are only examining warm-up characteristics, not ambient temperature changes, which can not only affect the VNA but the coaxial test cables as well.

Useful Radio Frequency Engineering Formulas and Charts
February 9, 2023
This application note is a collection of essential formulas and charts for Radio Frequency Engineering.

Introduction to the Metrology of VNA Measurement
May 31, 2022
A VNA measures reflection coefficients of RF devices and systems. To be very meticulous, each measurement should be accompanied by an uncertainty with a specified statistical confidence interval. To understand the measurement uncertainty, or accuracy, it is important to understand the contributing factors, which are explored in this application note.

Automation Guide for CMT VNAs
September 23, 2021
There are two methods for automating CMT VNAs. One is using SCPI commands and the other is using COM (Component Object Model) functions.