Moving in the table of calibration standard definitions (see figure) using navigation keys. Enter the parameter values for a calibration kit by using the navigation keys in the table of calibration standard definitions:
Standard Type |
Select the standard type: •Open •Short •Load •Thru •Data-Based |
Standard Label |
Standard labels specified on the calibration menu softkeys. |
C0·1e-15 F |
For an OPEN standard, C0 coefficient in the polynomial formula of the fringe capacitance:
|
C1·1e-27 F/Hz |
For an OPEN standard, C1 coefficient in the polynomial formula of the fringe capacitance. |
C2·1e-36 F/Hz2 |
For an OPEN standard, C2 coefficient in the polynomial formula of the fringe capacitance. |
C3·1e-45 F/Hz3 |
For an OPEN standard, C3 coefficient in the polynomial formula of the fringe capacitance. |
L0·1e-12 H |
For a SHORT standard, L0 coefficient in the polynomial formula of the residual inductance:
|
L1·1e-24 H/Hz |
For a SHORT standard, L1 coefficient in the polynomial formula of the residual inductance. |
L2·1e-33 H/Hz2 |
For a SHORT standard, L2 coefficient in the polynomial formula of the residual inductance. |
L3·1e-42 H/Hz3 |
For a SHORT standard, L3 coefficient in the polynomial formula of the residual inductance. |
Offset Delay, ps |
Offset delay value in one direction (ps). Can be switched to physical length (mm). The parameter is used only for the calibration standard model. |
Offset Z0, Ohm |
Offset characteristic impedance value (Ohm). For waveguide must be set to 1 Ohm. |
Offset Loss, GOhm/s |
Offset loss value (GOhm/s). The parameter is used only for the definition of the standard with the help of the calibration standard model. |
F min, MHz |
Minimum operating frequency of the coaxial standard. Lower cutoff frequency of the waveguide standard. |
F max, MHz |
Maximum operating frequency of the coaxial standard. Upper cutoff frequency of the waveguide standard. |
Media |
Coaxial or Waveguide |
H/W |
Waveguide height to width ratio. |
To set the type of a standard, double click on the field with the name of the standard in the Standard Type line and select the type in the opened Standard Type dialog: •Open or Short or Load or Thru — if the standard is described with an equivalent circuit. •Data-Dased — if the standard is described with the Table of S-parameters (See Calibration Standard Defining by S-Parameter File). To edit the label of a standard, double click on the field in the Standard Label line and enter a description for the standard using the on-screen keypad. To save the settings and close the dialog click Ok.
|
To edit the calibration standard parameters, and double click in the table on the corresponding cell. Enter the required value using the on-screen keypad. |
Parameters of a calibration standard can be set from an S-parameter file in Touchstone format (See Data-Based Calibration Standards).
To set the calibration standard parameters by S-parameter file double click in the table on the corresponding field in the Touchstone File line. The table of S-parameters can be filled downloaded from a file of Touchstone format. Files with *.S1P extension are used for one-port standards, and files with *.S2P extension are used for two-port standards.
|