SCPI Command

SENSe<Ch>:CORRection:COLLect:CKIT:STAN<Std>:C0 <numeric>

SENSe<Ch>:CORRection:COLLect:CKIT:STAN<Std>:C0?

Description

Sets or reads out the C0 value for the open calibration standard.

command/query

Target

Standard <Std> of the calibration kit specified for channel <Ch>,

<Ch>={[1]|2|3|4}

<Ch>={[1]…|16} (in N-port mode only)

<Std>={[1]|2|...K}, where K – the number of the standards in the calibration kit

Parameter

<numeric> the CO value from –1E18 to 1E18

Unit

1E–15 F (Farad)

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Equivalent Softkeys

Calibration > Calibration Kit > Edit Cal Kit > C0 [e–15 F]


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0

Syntax

Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0

app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1 = 100

Type

Double (read/write)


 

Back to SENSe