The trigger system operates at the analyzer level.
The Analyzer can be in one of the following three states:
•Stop — the Analyzer waits for the trigger system to enter the Waiting for Trigger state.
•Waiting for Trigger — measurement stops, the Analyzer waits for the trigger signal. If the Internal trigger source (see Trigger Source) is selected, it is automatically generated.
•Measurement Cycle — all channels are measured in turn.
The figure below shows the states of the Analyzer, and the transitions between them.
The table below describes the transitions between analyzer states.
Transition |
Condition |
Button |
Command |
---|---|---|---|
To Stop |
Power on |
— |
— |
Reset |
System > Preset |
||
Abort of the current measurement cycle. |
— |
||
Changing Analyzer settings by user or by the SCPI command. |
For example: Stimulus > Start |
For example: SENSe:FREQuency:STARt |
|
Stimulus > Trigger > Hold |
|||
Stop –> Waiting for Trigger |
Every time if the trigger initiation mode has been set to Continuous. |
Stimulus > Trigger > Continuous |
|
Once when the trigger initiation mode has been set to Single. |
Stimulus > Trigger > Single |
||
1.3 |
Automatically, if the trigger source is set to Internal. |
Stimulus > Trigger > Trigger source > Internal |
|
At a signal arrival at the external trigger input, if the trigger source is set to External. (except TR1300) |
Stimulus > Trigger > Trigger source > External |
||
Upon receipt of SCPI command, if the trigger source is set to Bus. |
Stimulus > Trigger > Trigger source > Bus |
||
1.4 Measurement Cycle –> Waiting for Trigger |
At the end of a measurement cycle, when the trigger initiation mode has been set to Continuous. |
Stimulus > Trigger > Continuous |
|
After measuring a point, when the On Point trigger function is active. (except TR1300) |
Stimulus > Trigger > Ext Trigger > Event > On Point |
— |
|
1.5 Measurement Cycle –> Stop |
At the end of a measurement cycle, when the Continuous trigger initiation mode is disabled. |
— |
— |