The trigger system operates at the analyzer level.
The Analyzer can be in one of the following three states:
•Stop — the Analyzer waits for the trigger system to enter the Waiting for Trigger state.
•Waiting for Trigger — measurement stops, the Analyzer waits for the trigger signal. If the Internal trigger source (see Trigger Source) is selected, it is automatically generated.
•Measurement Cycle — all channels are measured in turn.
The figure below shows the states of the Analyzer, and the transitions between them.
Analyzer States and Transitions
The table below describes the transitions between analyzer states.
Transition |
Condition |
Button |
Command |
---|---|---|---|
To Stop |
Power on |
— |
— |
Reset |
System > Preset |
||
Abort of the current measurement cycle. |
— |
||
Changing Analyzer settings by user or by the SCPI command. |
For example: Stimulus > Start |
For example: SENSe:FREQuency:STARt |
|
Stimulus > Trigger > Hold |
|||
Stop –> Waiting for Trigger |
Every time if the trigger initiation mode has been set to Continuous. |
Stimulus > Trigger > Continuous |
|
Once when the trigger initiation mode has been set to Single. |
Stimulus > Trigger > Single |
||
1.3 |
Automatically, if the trigger source is set to Internal. |
Stimulus > Trigger > Trigger source > Internal |
|
At a signal arrival at the external trigger input, if the trigger source is set to External. (except TR1300) |
Stimulus > Trigger > Trigger source > External |
||
Upon receipt of SCPI command, if the trigger source is set to Bus. |
Stimulus > Trigger > Trigger source > Bus |
||
1.4 Measurement Cycle –> Waiting for Trigger |
At the end of a measurement cycle, when the trigger initiation mode has been set to Continuous. |
Stimulus > Trigger > Continuous |
|
After measuring a point, when the On Point trigger function is active. (except TR1300) |
Stimulus > Trigger > Ext Trigger > Event > On Point |
— |
|
1.5 Measurement Cycle –> Stop |
At the end of a measurement cycle, when the Continuous trigger initiation mode is disabled. |
— |
— |