Please enable JavaScript to view this site.

 Network Analyzers using TRVNA software

SCPI Command

TRIGger[:SEQuence]:SINGle

Description

Generates a trigger signal and initiates a sweep under the following conditions.

Trigger source is set to the BUS (set by the command TRIG:SOUR BUS), otherwise an error occurs and the command is ignored.

Analyzer must be in the trigger waiting state, otherwise (the Analyzer is in the measurement state or in the hold state) an error occurs, and the command is ignored.

As opposed to the TRIG command this command is pending till the end of the sweep. The end of the sweep initiated by the TRIG:SING command can be waited using the *OPC? query.

no query

Related Commands

TRIG:SOUR

*OPC?

INIT:CONT

INIT

Equivalent Softkeys

None


Equivalent COM Command

SCPI.TRIGger.SEQuence.SINGle

Syntax

app.SCPI.TRIGger.SEQuence.SINGle

Type

Method


 

Back to TRIGger

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  24.1