Please enable JavaScript to view this site.

 Network Analyzers using TRVNA software

Reflection normalization is the simplest calibration method used for reflection coefficient measurements (S11). Measurement of one standard (SHORT or OPEN) is sufficient to perform this type of calibration (See figure below). This method is called normalization because the measured S-parameter at each frequency point is divided (normalized) by the corresponding S-parameter of the calibration standard. Reflection normalization corrects the reflection tracking error (Er) only. This constrains the accuracy of the method.


note

Reflection normalization can also be referred to as response open or response short calibration depending on the standard being used: OPEN or SHORT.


An optional LOAD standard measurement can be performed to correct the directivity error (Ed). The optional directivity calibration increases the accuracy of the reflection normalization.

Reflection normalization

Reflection Normalization

Before starting calibration perform the following settings: select active channel, set the parameters of the channel (frequency range, IF bandwidth, etc.), and select the calibration kit.

Response (open_short)

To open reflection normalization submenu, use the following softkeys:

Calibration > Calibrate > Response (Open) | Response (Short)

 

Response (open)

Connect an OPEN or a SHORT standard to the test port 1 as shown in above figure. Perform measurement using the Open or Short softkey respectively.

During the measurement, a pop up window will appear in the channel window. It will have Calibration label and will indicate the progress of the measurement.

On completion of the measurement, a checkmark will appear in the left part of the softkey.

Load (optional)

To perform the optional directivity calibration, connect a LOAD standard to the test port 1 as shown in the above figure and perform measurement using Load (Optional) softkey.

During the measurement, a pop up window will appear in the channel window. It will have Calibration label and will indicate the progress of the measurement. On completion of the measurement, a checkmark will appear in the left part of the softkey.

 

Apply

To complete the calibration procedure, click Apply.

This will activate the process of calibration coefficient table calculation and saving it into the memory. The error correction function will also be automatically enabled.

 

Cancel softkey

To clear the measurement results of the standards, click Cancel.

This softkey does not cancel the current calibration. To disable the current calibration turn off the error correction function (See Error Correction Disabling).

 

 

 


note

The calibration status can be checked in channel status bar (See General error correction status table) or in trace status field (See Trace error correction status table).


 

Rev.:  24.1