SCPI Command
SENSe<Ch>:CORRection:COLLect[:ACQuire]:LOAD <port>
Measures the calibration data of the load standard for the port 1. Note: The command starts the measurement for the channel independently of the trigger and trigger source settings. The command waits for the completion of the measurement. The command blocks the execution of the subsequent commands until the completion of the measurement. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...9}
<port> |
The number of the port is always 1 |
Calibration > Calibrate > Response (Open) > Load (Optional)
Calibration > Calibrate > Response (Short) > Load (Optional)
Calibration > Calibrate > Full 1–Port Cal > Load
Calibration > Calibrate > One Path 2–Port Cal > Load
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD = 1
Type
Long (write only)
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)