SENSe<Ch>:CORRection:COLLect:SAVE

Description

Calculates the calibration coefficients from the calibration standards measurements depending on the selected calibration type. The calibration type is selected by one of commands SENS:CORR:COLL:METH:XXXX.

On completion of the command, all the calibration standards measurements are cleared, and the error correction automatically turns ON.

At the attempt to execute this command before all the needed standards are measured, an error occurs, and the command is ignored.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...9}

Related Commands

Calibration type selection:

SENS:CORR:COLL:METH:OPEN

SENS:CORR:COLL:METH:SHOR

SENS:CORR:COLL:METH:THRU

SENS:CORR:COLL:METH:ERES

SENS:CORR:COLL:METH:SOLT1

Calibration standards measurement:

SENS:CORR:COLL:LOAD

SENS:CORR:COLL:OPEN

SENS:CORR:COLL:SHOR

SENS:CORR:COLL:THRU

Equivalent Softkeys

Calibration > Calibrate > {Response (Open) | Respose (Short) | Response (Thru) | Full 1-Port Cal | One Path 2-Port Cal} > Apply


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.COLLect.SAVE

Syntax

app.SCPI.SENSe(Ch).CORRection.COLLect.SAVE

Type

Method


 

Back to SENSe

Back to Command Finder