SCPI Command
SENSe<Ch>:CORRection:COLLect[:ACQuire]:SHORt <port>
Measures the calibration data of the short standard for the port 1. Note: The command starts the measurement for the channel independently of the trigger and trigger source settings. The command waits for the completion of the measurement. The command blocks the execution of the subsequent commands until the completion of the measurement. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...9}
<port> |
The number of the port is always 1 |
Error occurs. The command is ignored.
Calibration > Calibrate > Response (Short) > Short
Calibration > Calibrate > Full 1–Port Cal > Short
Calibration > Calibrate > One Path 2–Port Cal > Short
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.SHORt
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.SHORt = 1
Type
Long (write only)
Back to SENSe