SCPI Command
SENSe<Ch>:CORRection:COLLect:ECAL2:SAVE
Completes the procedure of the full 3-port or 4-port calibration that used 2-port ACM. The calibration type (3-port or 4-port) and port numbers are selected by commands SENS:CORR:COLL:ECAL2:METH:SOLT3 or SENS:CORR:COLL:ECAL2:METH:SOLT4. Calibration steps must first be performed using a 2-port ACM between different pairs of ports in such a way that there is at least one ACM connection to each port (see SENS:CORR:COLL:ECAL2). Optional THRU connections between other port pairs improve the calibration accuracy (see SENS:CORR:COLL:ECAL2:THRU). At the attempt to complete the calibration before all the needed steps are executed, an error occurs, and the command is ignored. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...16}
Calibration > AutoCal > 3-Port AutoCal > 2-Port AutoCal Module > Apply
Calibration > AutoCal > 4-Port AutoCal > 2-Port AutoCal Module > Apply
Equivalent COM Command
None