Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

SENSe<Ch>:CORRection:COLLect:ECAL2:METHod:SOLT3 <port1>,<port2>,<port3>

Description

Selects ports and sets the type to full 3-port for calibration performed with the 2-port AutoCal module. These presets are used on completion of the calibration executed by the SENS:CORR:COLL:ECAL2:SAVE command.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<port1>

The first port number from 1 to 4

<port2>

The second port number from 1 to 4

<port3>

The third port number from 1 to 4

Out of Range

If the same port numbers are specified, an error occurs.

Equivalent Softkeys

Calibration > AutoCal > 3-Port AutoCal > Select Ports {x-y-z}

Calibration > AutoCal > 3-Port AutoCal > 2-Port AutoCal Module


Equivalent COM Command

None


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 