SCPI Command
SENSe<Ch>:CORRection:COLLect:ECAL2:METHod:SOLT3 <port1>,<port2>,<port3>
Selects ports and sets the type to full 3-port for calibration performed with the 2-port AutoCal module. These presets are used on completion of the calibration executed by the SENS:CORR:COLL:ECAL2:SAVE command. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...16}
<port1> |
The first port number from 1 to 4 |
<port2> |
The second port number from 1 to 4 |
<port3> |
The third port number from 1 to 4 |
If the same port numbers are specified, an error occurs.
Calibration > AutoCal > 3-Port AutoCal > Select Ports {x-y-z}
Calibration > AutoCal > 3-Port AutoCal > 2-Port AutoCal Module
Equivalent COM Command
None
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)