 Network Analyzers using S4VNA software

SCPI Command

SENSe<Ch>:CORRection:COLLect[:ACQuire]:ECAL2:THRU <port 1>, <port 2>, {[UNKNown] | FLUSh}

Description

Measures a THRU between <port 1> and <port 2> when performing a full 3-port or 4-port calibration in the procedure that used 2-port ACM. This step is optional and can be performed for the port pairs that was not calibrated by the 2-port ACM. This step improves the calibration accuracy.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<port 1>

The number of the first port from 1 to 4

<port 2>

The number of the second port from 1 to 4

{[UNKNown] | FLUSh}

Selects type of the thru: unknown thru or flush thru (zero length thru). If not specified the unknown thru type is used.

Out of Range

If the same port numbers are specified, an error occurs.

Equivalent Softkeys

Calibration > AutoCal > 3-Port AutoCal > 2-Port AutoCal Module > n-m Thru (Optional) > {Flush Thru | Unknown Thru}

Calibration > AutoCal > 4-Port AutoCal > 2-Port AutoCal Module > Thru > n-m Thru >  {Flush Thru | Unknown Thru}


Equivalent COM Command

None


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 