SCPI Command
SENSe<Ch>:CORRection:COLLect[:ACQuire]:ECAL2:THRU <port 1>, <port 2>, {[UNKNown] | FLUSh}
Measures a THRU between <port 1> and <port 2> when performing a full 3-port or 4-port calibration in the procedure that used 2-port ACM. This step is optional and can be performed for the port pairs that was not calibrated by the 2-port ACM. This step improves the calibration accuracy. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...16}
<port 1> |
The number of the first port from 1 to 4 |
<port 2> |
The number of the second port from 1 to 4 |
{[UNKNown] | FLUSh} |
Selects type of the thru: unknown thru or flush thru (zero length thru). If not specified the unknown thru type is used. |
If the same port numbers are specified, an error occurs.
Calibration > AutoCal > 3-Port AutoCal > 2-Port AutoCal Module > n-m Thru (Optional) > {Flush Thru | Unknown Thru}
Calibration > AutoCal > 4-Port AutoCal > 2-Port AutoCal Module > Thru > n-m Thru > {Flush Thru | Unknown Thru}
Equivalent COM Command
None