SCPI Command

SENSe<Ch>:CORRection:COLLect:SIMPlified:SAVE

Description

Calculates the calibration coefficients for the simplified three- or four-port calibration from the calibration standards measurements when the three- or four-port calibration is selected as the calibration type. The calibration type is selected by one of commands SENS:CORR:COLL:METH:SOLT3/SOLT4/TRL3/TRL4.

The simplified three-port calibration allows omit one THRU measurement. The simplified four-port calibration allows omit up to three THRU measurements.

On completion of the command, all the calibration standards measurements are cleared and the error correction automatically turns ON.

If full set of calibration standard measurement is made this command behaves like the SENS:CORR:COLL:SAVE command.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Related Commands

Calibration type selection:

SENS:CORR:COLL:METH:SOLT3

SENS:CORR:COLL:METH:SOLT4

SENS:CORR:COLL:METH:TRL3

SENS:CORR:COLL:METH:TRL4

Calibration standards measurement:

SENS:CORR:COLL:ISOL

SENS:CORR:COLL:LOAD

SENS:CORR:COLL:OPEN

SENS:CORR:COLL:SHOR

SENS:CORR:COLL:THRU

SENS:CORR:COLL:TRLT

SENS:CORR:COLL:TRLL

SENS:CORR:COLL:TRLR

Equivalent Softkeys

Calibration > Calibrate > n-Port SOLT Cal | n-Port TRL Cal > Apply


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.COLLect.SIMPlified.SAVE

Syntax

app.SCPI.SENSe(Ch).CORRection.COLLect.SIMPlified.SAVE

Type

Method


 