SCPI Command
SENSe<Ch>:CORRection:COLLect:SIMPlified:SAVE
Calculates the calibration coefficients for the simplified three- or four-port calibration from the calibration standards measurements when the three- or four-port calibration is selected as the calibration type. The calibration type is selected by one of commands SENS:CORR:COLL:METH:SOLT3/SOLT4/TRL3/TRL4. The simplified three-port calibration allows omit one THRU measurement. The simplified four-port calibration allows omit up to three THRU measurements. On completion of the command, all the calibration standards measurements are cleared and the error correction automatically turns ON. If full set of calibration standard measurement is made this command behaves like the SENS:CORR:COLL:SAVE command. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...16}
Calibration type selection:
Calibration standards measurement:
Calibration > Calibrate > n-Port SOLT Cal | n-Port TRL Cal > Apply
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.SIMPlified.SAVE
app.SCPI.SENSe(Ch).CORRection.COLLect.SIMPlified.SAVE