SCPI Command
SENSe<Ch>:CORRection:COLLect[:ACQuire]:LOAD <port>
Measures the calibration data of the load standard for the specified port. Note: The command starts the measurement immediately if the trigger source for calibration set to the "Internal" by the command SENS:CORR:TRIG:FREE, otherwise waits for the trigger signal. The command blocks the execution of the subsequent commands until the completion of the measurement. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...16}
<port> |
The number of the port from 1 to 4 |
Calibration > Calibrate > Response (Open) > Load (Optional)
Calibration > Calibrate > Response (Short) > Load (Optional)
Calibration > Calibrate > 1–Port SOL Cal > Load
Calibration > Calibrate > One Path 2–Port Cal > Load
Calibration > Calibrate > 2–Port SOLT Cal > Port n Load
Calibration > Calibrate > 3–Port SOLT Cal > Reflection (Port n) > Load
Calibration > Calibrate > 4–Port SOLT Cal > Reflection (Port n) > Load
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD = 1
Type
Long (write only)
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)