Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

SENSe<Ch>:CORRection:COLLect[:ACQuire]:OPEN <port>

Description

Measures the calibration data of the open standard for the specified port.

Note: The command starts the measurement immediately if the trigger source for calibration set to the "Internal" by the command SENS:CORR:TRIG:FREE, otherwise waits for the trigger signal. The command blocks the execution of the subsequent commands until the completion of the measurement.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<port>

The number of the port from 1 to 4

Equivalent Softkeys

Calibration > Calibrate > Response (Open) > Open

Calibration > Calibrate > Full 1–Port Cal > Open

Calibration > Calibrate > One Path 2–Port Cal > Open

Calibration > Calibrate > 2–Port SOLT Cal > Port n Open

Calibration > Calibrate > 3–Port SOLT Cal > Reflection (Port n) > Open

Calibration > Calibrate > 4–Port SOLT Cal > Reflection (Port n) > Open


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN

Syntax

app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN = 1

Type

Long (write only)


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.4