The instrument status bar is located at the bottom of the screen.
Field Description |
Message |
Instrument Status |
Note |
---|---|---|---|
DSP status |
Not Ready |
No communication between DSP and PC. |
|
Loading |
DSP firmware is loading. |
||
Ready |
DSP is running normally. |
||
Sweep status |
Meas |
A sweep is in progress. |
For a detailed description, see Trigger Settings. |
Hold |
A sweep is on hold. |
||
Ext |
Waiting for External trigger. |
||
Man |
Waiting for Manual trigger. |
||
Bus |
Waiting for Bus trigger. |
||
Calibration |
Calibration… |
Calibration standard measurement is in progress. |
|
RF signal |
RF Off |
Stimulus signal output is turned OFF. |
For a detailed description, see RF Out Function. |
External reference frequency |
ExtRef |
External reference frequency input (10 MHz) is turned on. |
For a detailed description, see Reference Frequency Oscillator Selection. |
Display update |
Update Off |
Display update is turned OFF. |
For a detailed description, see Screen Update Setting. |
System correction status |
Sys Corr OFF |
System correction is turned OFF. |
For a detailed description, see System Correction Setting. |
Factory calibration error |
Factory Cal Failure: (PW, PT, LOP, ED, IF, VM, xPW, xPT) |
Calibration data in ROM has an error. The analyzer is damaged and requires factory repair. |
|
External power meter status |
Power Meter: (connection, connection error, ready, measurement, zero setting, zero setting error) |
Status of an external power meter when used with an Analyzer |
For a detailed description, see Power Meter Setting. |
Power Trip function |
Port n Power Trip at Overload |
Port n overload, stimulus signal output is turned OFF. |
For a detailed description, see Power Trip Function. |
Cycle Time |