SCPI Command
TRIGger[:SEQuence]:SINGle
Generates a trigger signal and initiates a sweep under the following conditions. •Trigger source is set to the BUS (set by the command TRIG:SOUR BUS), otherwise an error occurs and the command is ignored. •Analyzer must be in the trigger waiting state, otherwise (the Analyzer is in the measurement state or in the hold state) an error occurs, and the command is ignored. As opposed to the TRIG command this command is pending till the end of the sweep. The end of the sweep initiated by the TRIG:SING command can be waited using the *OPC? query. |
---|
no query |
None
Equivalent COM Command
SCPI.TRIGger.SEQuence.SINGle
app.SCPI.TRIGger.SEQuence.SINGle
Method
Back to TRIGger
Back to Command Finder
Back to Command Finder (From User Interface)