Removing fixture characteristics using the AFR plug-in involves following steps:
Provide general information about the test fixture and select a removal method. Recommendations for selecting the removal method can be found in this section.
Configure the VNA for optimal fixture removal results. VNA settings are displayed and reviewed in this step.
3.Calibrate the VNA
VNA calibration is essential for accurate fixture removal. VNA calibration status is displayed and reviewed in this step.
Connect each port of the fixture to the VNA and measure the calibration standards. The fixture removal correction is calculated and applied automatically to the VNA upon clicking the APPLY button in this step.
5.Save Correction Files and Setup (optional)
If desired, save the correction files and AFR setup for future use.