The channel status bar is located in the bottom part of the channel window (See figure below).

Channel Status Bar

Channel Status Bar

It contains the following elements:

Stimulus start field allows display and entry of the start frequency or power, depending on the sweep type. This field can be switched to indicate the stimulus center frequency, in this case the word «Start» will change to «Center». For a detailed description of stimulus setting, see Sweep Range.

Number of Points field allows display and entry of the number of sweep points. The number of points can be set from 2 to 16001. For a detailed description see Number of Points.

Sweep type field allows display and selection of the sweep type. The values of this field are represented in the table below. For a detailed description see Sweep Type.

Symbol

Definition

Lin

Linear frequency sweep

Log

Logarithmic frequency sweep

Seg

Segment frequency sweep

Pow

Power sweep

Voltmeter

Vector Voltmeter

IF bandwidth field allows display and setting of the IF bandwidth. The values can be set from the 10 Hz to 30 kHz. For a detailed description see IF Bandwidth Setting.

Power level field allows display and entry of the port output power. In power sweep mode, the field switches to indicate the CW frequency of the source. For a detailed description see Stimulus Power.

Stimulus stop field allows display and entry of the stop frequency or power, depending on the sweep type. This field can be switched to indication of stimulus span, in this case the word «Stop» will change to «Span». For a detailed description of stimulus setting, see Sweep Range.

Error correction field displays the integrated status of error correction for S-parameter traces and the integrated status of power correction for all the traces. The values of this field are represented in the table below. For a detailed description see Error Correction Status.

Symbol

Definition

--

No calibration data. No calibration was performed.

Cor

Error correction is enabled. The stimulus settings are the same for the measurement and the calibration.

C?

Error correction is enabled. The stimulus settings are not the same for the measurement and the calibration. Interpolation is applied.

C!

Error correction is enabled. The stimulus settings are not the same for the measurement and the calibration. Extrapolation is applied.

Off

Error correction is turned off.

PC

Power correction is enabled. The stimulus settings are the same for the measurement and the calibration.

PC?

Power correction is enabled. The stimulus settings are not the same for the measurement and the calibration. Interpolation is applied.

PC!

Power correction is enabled. The stimulus settings are not the same for the measurement and the calibration. Extrapolation is applied.

Averaging status field displays the averaging status, if this function is enabled. The first number is the averaging current counter value, the second one is the averaging factor. For a detailed description see Averaging Setting.

 