The trigger initiation mode determines how many trigger signals the Analyzer will accept. The trigger system can operate in one of the following three modes (See the table below).
Trigger Initiation Mode |
Function |
---|---|
Continuous [default] |
The Analyzer accepts an infinite number of trigger signals, the measurement cycle is executed every time a trigger signal is detected. The measurement cycle includes sequential measurement of all opened channels one after another. |
Single |
The Analyzer accepts one trigger signal. One measurement cycle is executed at one trigger signal detection after the mode has been enabled. The measurement cycle includes sequential measurement of all opened channels one after another.The trigger system turns to Stop state after the measurement cycle is complete. |
Hold |
The Analyzer accepts no trigger signals. The trigger system is in Stop state. |
To set the trigger initiation mode, use the following softkeys: Stimulus > Trigger Then select the required trigger initiation mode: •Hold •Single •Continuous |
|
|