Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<ch>:PLS2:ADVanced:PGEN<pg>:EDGE {POS|NEG}

SENSe<ch>:PLS2:ADVanced:PGEN<pg>:EDGE?

Description

Sets the active edge of the trigger to start the pulse generator PG1...PG6.

Note: The command is used in advanced mode, it is not used in basic pulse modes. Application requires knowledge of the internal structure of the analyzer.

Note: S5243 model only.

command/query

Target

Channel <Ch> = {[1]|2| ... 16},

Generator <pg> = {1|2|3|4|5|6|7}

Parameter

<char> active edge of the trigger:

POS

Positive (rising edge)

NEG

Negative (falling edge)

Query Response

{POS|NEG}

Preset Value

POS

Related Commands

SENS:PLS2:ADV:APPL

Equivalent Softkeys

Stimulus > Pulse Meas > Advanced Settings... -> Pulse Generators -> Edge -> {POS | NEG}


Equivalent COM Command

None


 

Back to Pulse Measurement S5243

Back to Command Finder (From User Interface)

 

Rev.:  24.1