SCPI Command
SENSe<ch>:PLS2:ADVanced:PGEN<pg>:COUNt <numeric>
SENSe<ch>:PLS2:ADVanced:PGEN<pg>:COUNt?
Sets the N — number of pulses in a burst for pulse generator PG1...PG6 operating in BURSt mode. For more details, see the figure Pulse generator timing diagram. Note: The command is used in advanced mode, it is not used in basic pulse modes. Application requires knowledge of the internal structure of the analyzer. Note: S5243 model only. |
---|
command/query |
Channel <Ch> = {[1]|2| ... 16},
Generator <pg> = {1|2|3|4|5|6|7}
<numeric> number of pulses in a burst.
<numeric>
0
Stimulus > Pulse Meas > Advanced Settings... -> Pulse Generators -> Width2/Cnt
Equivalent COM Command
None
Back to Pulse Measurement S5243