Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<ch>:PLS2:ADVanced:PGEN<pg>:COUNt <numeric>

SENSe<ch>:PLS2:ADVanced:PGEN<pg>:COUNt?

Description

Sets the N — number of pulses in a burst for pulse generator PG1...PG6 operating in BURSt mode. For more details, see the figure Pulse generator timing diagram.

Note: The command is used in advanced mode, it is not used in basic pulse modes. Application requires knowledge of the internal structure of the analyzer.

Note: S5243 model only.

command/query

Target

Channel <Ch> = {[1]|2| ... 16},

Generator <pg> = {1|2|3|4|5|6|7}

Parameter

<numeric> number of pulses in a burst.

Query Response

<numeric>

Preset Value

0

Related Commands

SENS:PLS2:ADV:APPL

Equivalent Softkeys

Stimulus > Pulse Meas > Advanced Settings... -> Pulse Generators -> Width2/Cnt


Equivalent COM Command

None


 

Back to Pulse Measurement S5243

Back to Command Finder (From User Interface)

 

Rev.:  24.1