Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<ch>:PLS2:ADVanced:LINK:SYNC1 {EXT1...EXT6|RDY}

SENSe<ch>:PLS2:ADVanced:LINK:SYNC1?

Description

Sets the signal source for the Sync1 line. The signal source is determined by the command parameter.

Note: The command is used in advanced mode, it is not used in basic pulse modes. Application requires knowledge of the internal structure of the analyzer.

Note: S5243 model only.

command/query

Target

Channel <Ch> = {[1]|2| ... 16}

Parameter

<char> the signal source for the Sync1 line:

EXT1

"Ext Trig Out" signal

EXT2

"Ext Trig In" signal

EXT3

"Pulse Trig 3" signal

EXT4

"Pulse Trig 4" signal

EXT5

"Pulse Trig 5" signal

EXT6

"Pulse Trig 6" signal

RDY

Ready (ready for measurement, frequency set)

Query Response

{EXT1...EXT6|RDY}

Related Commands

SENS:PLS2:ADV:APPL

Equivalent Softkeys

Stimulus > Pulse Meas > Advanced Settings... > Sync 1 -> {source}


Equivalent COM Command

None


 

Back to Pulse Measurement S5243

Back to Command Finder (From User Interface)

 

Rev.:  24.1