SCPI Command
SENSe<ch>:PLS2:ADVanced:CONNector<n>:OUTput:ROUTe {PGEN1...PGEN7|EXT1...EXT6|SYNC1|SYNC2|LOW|HIGH|STRG}
SENSe<ch>:PLS2:ADVanced:CONNector<n>:OUTput:ROUTe?
Sets the signal source for the specified analyzer connector <n> if the signal direction at that connector is set to OUT with the SENS:PLS2:ADV:CONN:DIR command. The signal source is determined by the command parameter. Note: The command is used in advanced mode, it is not used in basic pulse modes. Application requires knowledge of the internal structure of the analyzer. Note: S5243 model only. |
---|
command/query |
Channel <Ch> = {[1]|2| ... 16},
Connector <n>:
1 |
"Ext Trig Out" |
2 |
"Ext Trig In" |
3 |
"Pulse Trig 3" |
4 |
"Pulse Trig 4" |
5 |
"Pulse Trig 5" |
6 |
"Pulse Trig 6" signal |
Sets the signal source for the connector <n>:
PGEN1...PGEN7 |
Signal from pulse generator PG1...PG7 |
EXT1 |
Input signal at Ext Trig Out connector |
EXT2 |
Input signal at Ext Trig In connector |
EXT3 |
Input signal at Pulse Trig 3 connector |
EXT4 |
Input signal at Pulse Trig 4 connector |
EXT5 |
Input signal at Pulse Trig 5 connector |
EXT6 |
Input signal at Pulse Trig 6 connector |
SYNC1 |
Internal line SYNC1 |
SYNC2 |
Internal line SYNC2 |
LOW |
Constant low level |
HIGH |
Constant high level |
STRG |
Trigger output (in standard mode) |
{PGEN1...PGEN7|EXT1...EXT6|SYNC1|SYNC2|LOW|HIGH|STRG}
Stimulus > Pulse Meas > Advanced Settings... > {Ext Trig Out | Ext Trig In | Pulse Trig 3 | Pulse Trig 4 | Pulse Trig 5 | Pulse Trig 6} -> Output: {source}
Equivalent COM Command
None
Back to Pulse Measurement S5243