Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<ch>:PLS2:ADVanced:CONNector<n>:OUTput:ROUTe {PGEN1...PGEN7|EXT1...EXT6|SYNC1|SYNC2|LOW|HIGH|STRG}

SENSe<ch>:PLS2:ADVanced:CONNector<n>:OUTput:ROUTe?

Description

Sets the signal source for the specified analyzer connector <n> if the signal direction at that connector is set to OUT with the SENS:PLS2:ADV:CONN:DIR command. The signal source is determined by the command parameter.

Note: The command is used in advanced mode, it is not used in basic pulse modes. Application requires knowledge of the internal structure of the analyzer.

Note: S5243 model only.

command/query

Target

Channel <Ch> = {[1]|2| ... 16},

Connector <n>:

1

"Ext Trig Out"

2

"Ext Trig In"

3

"Pulse Trig 3"

4

"Pulse Trig 4"

5

"Pulse Trig 5"

6

"Pulse Trig 6" signal

Parameter

Sets the signal source for the connector <n>:

PGEN1...PGEN7

Signal from pulse generator PG1...PG7

EXT1

Input signal at Ext Trig Out connector

EXT2

Input signal at Ext Trig In connector

EXT3

Input signal at Pulse Trig 3 connector

EXT4

Input signal at Pulse Trig 4 connector

EXT5

Input signal at Pulse Trig 5 connector

EXT6

Input signal at Pulse Trig 6 connector

SYNC1

Internal line SYNC1

SYNC2

Internal line SYNC2

LOW

Constant low level

HIGH

Constant high level

STRG

Trigger output (in standard mode)

Query Response

{PGEN1...PGEN7|EXT1...EXT6|SYNC1|SYNC2|LOW|HIGH|STRG}

Related Commands

SENS:PLS2:ADV:APPL

Equivalent Softkeys

Stimulus > Pulse Meas > Advanced Settings... > {Ext Trig Out | Ext Trig In | Pulse Trig 3 | Pulse Trig 4 | Pulse Trig 5 | Pulse Trig 6} -> Output: {source}  


Equivalent COM Command

None


 

Back to Pulse Measurement S5243

Back to Command Finder (From User Interface)

 

Rev.:  24.1