Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<ch>:PLS2:ADVanced:CONNector<n>:DIRection {INP|OUTP}

SENSe<ch>:PLS2:ADVanced:CONNector<n>:DIRection?

Description

Sets the signal direction (IN or OUT) for the specified connector on the analyzer's rear panel.

Note: The command is used in advanced mode, it is not used in basic pulse modes. Application requires knowledge of the internal structure of the analyzer.

Note: S5243 model only.

command/query

Target

Channel <Ch> = {[1]|2| ... 16};

Connector <n>:

1

"Ext Trig Out"

2

"Ext Trig In"

3

"Pulse Trig 3"

4

"Pulse Trig 4"

5

"Pulse Trig 5"

6

"Pulse Trig 6"

Parameter

<char> signal direction for the specified connector:

INP

Input

OUTP

Output

Query Response

{INP|OUTP}

Related Commands

SENS:PLS2:ADV:APPL

Equivalent Softkeys

Stimulus > Pulse Meas > Advanced Settings... > {Ext Trig Out | Ext Trig In | Pulse Trig 3 | Pulse Trig 4 | Pulse Trig 5 | Pulse Trig 6} -> {Input | Output}


Equivalent COM Command

None


 

Back to Pulse Measurement S5243

Back to Command Finder (From User Interface)

 