SCPI Command
SENSe<ch>:PLS2:ADVanced:LINK:GATing {PGEN1...PGEN7|LOW|HIGH}
SENSe<ch>:PLS2:ADVanced:LINK:GATing?
Sets the signal source at the input of the software gating block. A high signal level passes the ADC data unchanged, a low signal level resets the ADC data. Note: The command is used in advanced mode, it is not used in basic pulse modes. Application requires knowledge of the internal structure of the analyzer. Note: S5243 model only. |
---|
command/query |
Channel <Ch> = {[1]|2| ... 16}
<char> software gating source:
PGEN1...PGEN7 |
Generator PG1...PG7 |
LOW |
Constant low level |
HIGH |
Constant high level |
{PGEN1...PGEN7|LOW|HIGH}
Stimulus > Pulse Meas > Advanced Settings... > SW Gating -> {source}
Equivalent COM Command
None
Back to Pulse Measurement S5243
Back to Command Finder (From User Interface)