SCPI Command
SENSe<ch>:PLS2:ADVanced:LINK:FMOD {PGEN1...PGEN7|EXT1...EXT6|LOW|HIGH|SYNC2}
SENSe<ch>:PLS2:ADVanced:LINK:FMOD?
Sets the signal source at the input of the fast modulator. The signal source is determined by the command parameter. A high signal level opens the pulse modulator. Note: The command is used in advanced mode, it is not used in basic pulse modes. Application requires knowledge of the internal structure of the analyzer. Note: S5243 model only. |
---|
command/query |
Channel <Ch> = {[1]|2| ... 16}
<char> signal source at the input of the fast modulator:
PGEN1...PGEN7 |
Generator PG1...PG7 |
EXT1 |
"Ext Trig Out" signal |
EXT2 |
"Ext Trig In" signal |
EXT3 |
"Pulse Trig 3" signal |
EXT4 |
"Pulse Trig 4" signal |
EXT5 |
"Pulse Trig 5" signal |
EXT6 |
"Pulse Trig 6" signal |
SYNC2 |
Internal line SYNC2 signal |
LOW |
Constant low level |
HIGH |
Constant high level |
{PGEN1...PGEN7|EXT1...EXT6|LOW|HIGH|SYNC2}
Stimulus > Pulse Meas > Advanced Settings... > Fast Modulator -> {source}
Equivalent COM Command
None
Back to Pulse Measurement S5243
Back to Command Finder (From User Interface)