SCPI Command
SENSe<Ch>:CORRection:COLLect[:ACQuire]:SHORt <port>
Measures the calibration data of the short standard for the specified port. Note: The command starts the measurement for the channel independently of the trigger and trigger source settings. The command waits for the completion of the measurement. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|3|4}
<Ch>={[1]…|16} (in N-port mode only)
Port number is 1,
1-16 (in N-port mode only).
Calibration > Short
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.SHORt
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.SHORt = Port
Type
Long (write only)
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)