SCPI Command

SENSe<Ch>:CORRection:COLLect:METHod:SOLT2 <numeric1>,<numeric2>

Description

Selects the port and sets the full two-port (SOLT) calibration type for the calculation of the calibration coefficients on completion of the calibration executed by the SENS:CORR:COLL:SAVE command.

(in N-port mode only)

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16} (in N-port mode only)

Parameter

Port number is 1,

1-16 (in N-port mode only)

Out of Range

If the same port numbers are specified, an error occurs.

Related Commands

SENS:CORR:COLL:SAVE

Equivalent Softkeys

Calibration > Calibration type > Full 2-Port Cal with Scalar Thru


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT2

Syntax

app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT2 = Array(2, 1)


NOTE

The array contains 2 elements:

Data(0) is the number of the receiver port.

Data(1) is the number of the source port.


Type

Variant (array of Long) (write only)


 