SCPI Command
AFR:SYSTem:STEP<Step>:THRU:OFFSet {OFF|ON|0|1}
Sets the manual input mode of the calibration plane offset at the <Step> measurement step. 2xThrough method. To set the offset values of each fixture, use the AFR:CALC:STEP<Step>:THRU:OFFS <numeric 1> , <numeric 2> command. The command disables automatic determination mode of the calibration plane offset if it was previously enabled by the AFR:SYST:STEP<Step>:THRU:REFOFFS 1 command. Execution of the command is ignored if the measurements were not performed at step <Step>. To find out the state of the measurement step, use the AFR:SYST:STEP<Step>:MEAS? command. |
---|
no query |
Measurement step <Step>,
<Step> = {1|2|...}
{ON|1} |
ON |
{OFF|0} |
OFF |
AFR:CALCulate:STEP<Step>:THRU:OFFSet <numeric 1>,<numeric 2>
AFR: SYST:STEP<Step>:MEASured?
Fixture measurement > Step <Step> of N > 2xThrough method > Apply fixture F1/F2 calibration plane offset
F1 - name of the left fixture,
F2 - name of the right fixture,
N - maximum number of steps.