Automatic Fixture Removal

SCPI Command

AFR:SYSTem:STEP<Step>:THRU:OFFSet {OFF|ON|0|1}

Description

Sets the manual input mode of the calibration plane offset at the <Step> measurement step.

2xThrough method. To set the offset values of each fixture, use the AFR:CALC:STEP<Step>:THRU:OFFS <numeric 1> , <numeric 2> command.

The command disables automatic determination mode of the calibration plane offset if it was previously enabled by the AFR:SYST:STEP<Step>:THRU:REFOFFS 1 command.

Execution of the command is ignored if the measurements were not performed at step <Step>. To find out the state of the measurement step, use the AFR:SYST:STEP<Step>:MEAS? command.

no query

Target

Measurement step <Step>,

<Step> = {1|2|...}

Parameter

{ON|1}

ON

{OFF|0}

OFF

Related Commands

AFR:CALCulate:STEP<Step>:THRU:OFFSet <numeric 1>,<numeric 2>

AFR: SYST:STEP<Step>:MEASured?

Related Controls of User Interface

Fixture measurement > Step <Step> of N > 2xThrough method > Apply fixture F1/F2 calibration plane offset

F1 - name of the left fixture,

F2 - name of the right fixture,

N - maximum number of steps.