SCPI Command

AFR:CALCulate:STEP<Step>:THRU:OFFset <numeric 1>,<numeric 2>

Description

2xThrough method.

Sets the offset values of the fixture calibration plane at the <Step> measurement step.

First, activate the calibration plane offset mode by the AFR:SYST:STEP<Step>:THRU:OFFS 1 command.

no query

Target

Offset value of the calibration plane at the measurement step <Step>,

<Step> = {1|2|...}

Parameter

<numeric 1>

Left fixture offset, is specified in picoseconds

<numeric 2>

Right  fixture offset, is specified in picoseconds

Related Commands

AFR:SYSTem:STEP<Step>:THRU:OFFSet

Related Controls of User Interface

Fixture measurement > Step <Step> of N > 2xThrough method > Apply fixture F1/F2 calibration plane offset,

F1 - name of the left fixture,

F2 - name of the right fixture,

N - maximum number of steps.

Back to AFR:CALCulate

 