Please enable JavaScript to view this site.

Automatic Fixture Removal

SCPI command

AFR:CALCulate:STEP<Step>:REFlect:OFFSet <numeric>

Описание

1xReflect method.

Sets the offset value of the fixture calibration plane in the <Step> measurement step.

First, activate the calibration plane offset mode by the AFR:SYST:STEP<Step>:REF:OFFS 1 command.

no query

Target

Value of the calibration plane offset at the <Step> measurement step,

<Step> = {1|2|...}

Parameter

<numeric> offset value, is specified in picoseconds

Related commands

SYSTem:STEP<Step>:REFlect:OFFSet

Related Controls of User Interface

Fixture measurement > Step <Step> из N  > Apply fixture calibration plane offset

N - maximum number of steps.

Back to AFR:CALCulate

 

Rev.:  22.2.1