Please enable JavaScript to view this site.

Automatic Fixture Removal

SCPI Command

AFR:CALCulate:STEP<Step>:DUT

Description

Performs input and output reflection measurement of the fixture with the DUT. Measurement results are used to correct the resultant S-parameters of the fixture, specifically the reflection coefficients. Other S-parameters remain unchanged from the previous measurement steps, including the entered calibration plane offset values.

no query

Target

Measurement step <Step> of the fixture with the DUT,

<Step> = {1|2|...}

Related Controls of User Interface

Fixture measurement > Step <Step> of N  > Fixture with DUT > Measure

N - maximum number of steps.

Back to AFR:CALCulate

 

Rev.:  22.2.1