SCPI Command
SENSe:CORRection:COLLect:CKIT:STAN<Std>:LOSS <numeric>
SENSe:CORRection:COLLect:CKIT:STAN<Std>:LOSS?
Sets or reads out the offset loss value for the calibration standard. |
---|
command/query |
Standard <Std> of the calibration kit,
<Std>={[1]|2|...N}, where N is the number of the standards in the calibration kit
<numeric> the offset loss value from –1E18 to 1E18
Ω/s (Ohm/second)
Sets the value of the limit, which is closer to the specified value.
<numeric>
Calibration > Cal Kit > Edit Cal Kit > Offset Loss
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LOSS
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LOSS
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LOSS = 700E6
Double (read/write)
Back to SENSe