SCPI Command
SENSe:CORRection:COLLect:CKIT:STAN<Std>:C1 <numeric>
SENSe:CORRection:COLLect:CKIT:STAN<Std>:C1?
Sets or reads out the C1 value for the open calibration standard. |
---|
command/query |
Standard <Std> of the calibration kit,
<Std>={[1]|2|...N}, where N is the number of the standards in the calibration kit
<numeric> the C1 value from –1E18 to 1E18
1E–27 F/Hz (Farad/Hertz)
Out of Range
Sets the value of the limit, which is closer to the specified value.
<numeric>
Calibration > Cal Kit > Edit Cal Kit > Capacitance > C1 10–27 F/Hz
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0 = 100
Double (read/write)
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)