SCPI Command
SENSe:CORRection:COLLect:CKIT:STAN<Std>:C0 <numeric>
SENSe:CORRection:COLLect:CKIT:STAN<Std>:C0?
Sets or reads out the C0 value for the open calibration standard. |
---|
command/query |
Standard <Std> of the calibration kit,
<Std>={[1]|2|...N}, where N is the number of the standards in the calibration kit
<numeric> the C0 value from –1E18 to 1E18
1E–15 F (Farad)
Out of Range
Sets the value of the limit, which is closer to the specified value.
<numeric>
Calibration > Cal Kit > Edit Cal Kit > Capacitance > C0 10–15 F
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1 = 100
Double (read/write)
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)