SCPI Command
SENSe<Ch>:CORRection:COLLect:THRU:ADDition:FULL4:ACQuire <port1>,<port2>
Measures an unknown thru between the specified ports. The measurements are used to complete the four-port thru addition function by the command SENS:CORR:COLL:THRU:ADD:FULL4:COMP. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...16}
<port1> |
The first port number from 1 to 4 |
<port2> |
The second port number from 1 to 4 |
Calibration > Calibrate > Thru Addition > Comlete 4-Port Calibration > Thru {m-n}
Equivalent COM Command
None
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)