SCPI Command

SENSe<Ch>:CORRection:COLLect:THRU:ADDition:FULL3:COMPlete

Description

Completes the full three-port calibration between the ports specified by the command SENS:CORR:COLL:THRU:ADD:FULL3:PORT. The ports must be calibrated using the full one-port calibration in advance. The necessary number of the thru measurement must be accomplished by the command SENS:CORR:COLL:THRU:ADD:FULL3:ACQ.

This command calculates the error terms Et and El using the unknown thru algorithm. Then it saves the Et and El error terms to the existing calibration getting the full three-port calibration from the three one-port calibrations.

If the full three-port calibration already existed, updates the Et and El error terms.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Equivalent Softkeys

Calibration > Calibrate > Thru Addition > Comlete 3-Port Calibration > Apply


Equivalent COM Command

None


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 