Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

SENSe<Ch>:CORRection:COLLect:THRU:ADDition:FULL3:ACQuire <port1>,<port2>

Description

Measures an unknown thru between the specified ports. The measurements are used to complete the three-port calibration in the thru addition function by the command SENS:CORR:COLL:THRU:ADD:FULL3:COMP.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<port1>

The first port number from 1 to 4

<port2>

The second port number from 1 to 4

Equivalent Softkeys

Calibration > Calibrate > Thru Addition > Comlete 3-Port Calibration > Thru {m-n}


Equivalent COM Command

None


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.4