The Analyzer supports several methods of calibration. The calibration methods vary by quantity and type of the standards being used, by type of error correction, and accuracy. The table below presents an overview of calibration methods.
Calibration Method |
Parameters |
Standards |
Errors |
Accuracy |
---|---|---|---|---|
S11 or S22 |
•SHORT or OPEN •LOAD |
Er1, Ed11 or Er2, Ed21 |
Low |
|
S21 or S12 |
•THRU •2 LOADs |
Et1, Ex12 or Et2, Ex22 |
Low |
|
S11 or S22 |
•SHORT •OPEN •LOAD |
Er1, Ed1, Es1 or Er2, Ed2, Es1 |
High |
|
S11, S21 or S12, S22 |
•SHORT •OPEN •LOAD •THRU •2 LOADs |
Er1, Ed1, Es1, Et1, Ex12 or Er2, Ed2, Es2, Et2, Ex22 |
Medium |
|
S11, S21 S12, S22 |
•SHORT •OPEN •LOAD •THRU •2 LOADs |
Er1, Ed1, Es1, Et1, El1, Ex12 Er2, Ed2, Es2, Et2, El2, Ex22 |
High |
|
S11, S21, S31 S12, S22, S32 S13, S23, S33 |
•SHORT •OPEN •LOAD •THRU •2 LOADs |
Twenty seven error terms (See table) |
High |
|
S11, S21, S31, S41 S12, S22, S32, S42 S13, S23, S33, S43 S14, S24, S34, S44 |
•SHORT •OPEN •LOAD •THRU •2 LOADs |
Forty eight error terms (See table) |
High |
|
S11, S21 S12, S22 |
•THRU or LINE •REFLECT •LINE or 2 LOADs |
Er1, Ed1, Es1, Et1, El1 Er2, Ed2, Es2, Et2, El2 |
Very High |
|
S11, S21, S31 S12, S22, S32 S13, S23, S33 |
•THRU or LINE • REFLECT •LINE or 2 LOADs |
The same as full free-port calibration except for isolation terms |
Very High |
|
S11, S21, S31, S41 S12, S22, S32, S42 S13, S23, S33, S43 S14, S24, S34, S44 |
•THRU or LINE • REFLECT •LINE or 2 LOADs |
The same as full four-port calibration except for isolation terms |
Very High |
|