Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<ch>:PLS2:MEASurement:DELay <numeric>

SENSe<ch>:PLS2:MEASurement:DELay?

Description

Sets the measurement start delay relative to the pulse trigger. Used in the “Point in pulse”, “Pulse to pulse” modes to align the measurement position within the modulation pulse.

Note: The modulated RF pulse has a delay relative to the start of the measurement due to the RF signal delay in the modulator and ADC. The typical value of this delay is 400 ns for a fast modulator, and 1.2 µs for a normal modulator. It is recommended to set the measurement delay to at least the specified value to compensate for the delay in the modulator and ADC.

Note: S5243 model only.

command/query

Target

Channel <Ch> = {[1]|2| ... 16}

Parameter

<numeric> measurement start delay value.

Query Response

<numeric>

Unit

sec (second)

Resolution

20 ns

Preset Value

400 ns

Related Commands

SENS:PLS2:MEAS:WIDT

Equivalent Softkeys

Stimulus > Pulse Meas > Meas Delay


Equivalent COM Command

None


 

Back to Pulse Measurement S5243

Back to Command Finder (From User Interface)

 

Rev.:  24.1