SCPI Command
SENSe<ch>:PLS2:MEASurement:DELay <numeric>
SENSe<ch>:PLS2:MEASurement:DELay?
Sets the measurement start delay relative to the pulse trigger. Used in the “Point in pulse”, “Pulse to pulse” modes to align the measurement position within the modulation pulse. Note: The modulated RF pulse has a delay relative to the start of the measurement due to the RF signal delay in the modulator and ADC. The typical value of this delay is 400 ns for a fast modulator, and 1.2 µs for a normal modulator. It is recommended to set the measurement delay to at least the specified value to compensate for the delay in the modulator and ADC. Note: S5243 model only. |
---|
command/query |
Channel <Ch> = {[1]|2| ... 16}
<numeric> measurement start delay value.
Query Response
<numeric>
sec (second)
Resolution
20 ns
400 ns
Stimulus > Pulse Meas > Meas Delay
Equivalent COM Command
None
Back to Pulse Measurement S5243
Back to Command Finder (From User Interface)