 Network Analyzers using S2VNA software

SCPI Command

SENSe<ch>:PLS2:MEASurement:WIDTh <numeric>

SENSe<ch>:PLS2:MEASurement:WIDTh?

Description

Sets the measurement width, which is equal to the measurement width of one of the analyzer's digital IF filters. When setting an arbitrary value, the closest discrete value corresponding to the IF bandwidth is selected. Used in the “Point in Pulse”, “Pulse to Pulse”, “Narrowband” modes.

Warning! The measurement width should not exceed the width of the modulation pulse in the wideband detection modes "Point in Pulse" and "Pulse to Pulse".  The measurement must be performed within a single modulation pulse.

In the "Narrowband" mode, the measurement width should be at least the width of a burst of N pulses, so that the energy of these pulses accumulates over the measurement width of the IF digital filter. It is recommended to set the value of N to at least 10.

Note: S5243 model only.

command/query

Target

Channel <Ch> = {[1]|2| ... 16}

Parameter

<numeric> measurement width value.

Unit

sec (second)

Query Response

<numeric>

Preset Value

In wideband detection modes 5.04 µs

In narrowband mode 1.01 ms

Related Commands

SENS:PLS2:MEAS:DEL

Equivalent Softkeys

Stimulus > Pulse Meas > Meas Width


Equivalent COM Command

None


 