SCPI Command
SENSe<ch>:PLS2:ADVanced:PGEN<pg>:MODE {OFF|1PULSe|2PULSes|3PULSes|BURSt|CONTinous}
SENSe<ch>:PLS2:ADVanced:PGEN<pg>:MODE?
Sets the pulse generator mode PG1...PG6. For more details, see the figure Pulse generator timing diagram. Note: The command is used in advanced mode, it is not used in basic pulse modes. Application requires knowledge of the internal structure of the analyzer. Note: S5243 model only. |
---|
command/query |
Channel <Ch> = {[1]|2| ... 16},
Generator <pg> = {1|2|3|4|5|6|7}
<char> generation mode:
OFF |
Generation disabled |
1PULSe |
Template 1 pulse |
2PULSes |
Template 2 pulses |
3PULSes |
Template 3 pulses |
BURSt |
Generation of N pulses in the burst |
CONTinous |
Continuous generation |
{OFF|1PULSe|2PULSes|3PULSes|BURSt|CONTinous}
Stimulus > Pulse Meas > Advanced Settings... -> Pulse Generators -> Mode
Equivalent COM Command
None
Back to Pulse Measurement S5243
Back to Command Finder (From User Interface)