Please enable JavaScript to view this site.

 Network Analyzers using RVNA and RNVNA software

SCPI Command

SENSe<Ch>:FREQuency:CENTer <frequency>

SENSe<Ch>:FREQuency:CENTer?

Description

Sets or reads out the stimulus center value of the sweep range for linear or logarithmic sweep type.

command/query

Target

Channel <Ch>,

<Ch>={[1]|2|3|4}

<Ch>={[1]…|16} (in N-port mode only)

Parameter

<frequency> the stimulus center value within the frequency limits of the analyzer.

Unit

Hz (Hertz)

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Preset Value

The center frequency of the Analyzer.

Equivalent Softkeys

Stimulus > Center


Equivalent COM Command

SCPI.SENSe(Ch).FREQuency.CENTer

Syntax

Value = app.SCPI.SENSe(Ch).FREQuency.CENTer

app.SCPI.SENSe(Ch).FREQuency.CENTer = 1E9

Type

Double (read/write)


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  24.1