Please enable JavaScript to view this site.

 Network Analyzers using RVNA and RNVNA software

SCPI Command

SENSe<Ch>:CORRection:EXTension:AUTO:CONFig {CSPN|AMKR|USPN}

SENSe<Ch>:CORRection:EXTension:AUTO:CONFig?

Description

Specifies the frequency range used for calculation of the results of the Auto Port Extension function.

command/query

Target

Channel <Ch>,

<Ch>={[1]|2|3|4}

<Ch>={[1]|2|...16} (in N-port mode only)

Parameter

CSPN

Uses current frequency span.

AMKR

Uses the frequency of the active marker. This is applied to Loss 1 and Loss 2 is ignored.

USPN

Uses arbitrary frequency range.

Query Response

{CSPN|AMKR|USPN}

Preset Value

CSPN

Equivalent Softkeys

Calibration > Port Extension > Auto Port Extension > Method {Current Span | Acive Marker | User Span}


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.EXTension.AUTO.CONFig

Syntax

Value = app.SCPI.SENSe(Ch).CORRection.EXTension.AUTO.CONFig = "CSPN"

Type

String (read/write)


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  24.1