Please enable JavaScript to view this site.

 Network Analyzers using RVNA and RNVNA software

SCPI Command

SENSe<Ch>:CORRection:COLLect:DATA:LOAD <port>,<numeric list>

SENSe<Ch>:CORRection:COLLect:DATA:LOAD? <port>

Description

Writes or reads out the array of the load calibration standard measurement for the port <port>.

The array size is 2N, where N is the number of measurement points.

For the n–th point, where n from 1 to N:

<numeric 2n–1> real part of the measurement

<numeric 2n> imaginary part of the measurement

(1-port VNA only)

command/query

Target

Channel <Ch>,

<Ch>={[1]|2|3|4}

Parameter

<port>

The port number 1

<numeric list>

The data array of the load standard measurement

Query Response

<numeric 1>, <numeric 2>, …<numeric 2N>

Related Commands

SENS:CORR:COLL:LOAD

Equivalent Softkeys

None


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.COLLect.DATA.LOAD(Pt)

Syntax

app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.LOAD (Pt)

app.SCPI.SENSe(Ch).CORRection.COLLect.DATA.LOAD (Pt) = Data

Type

Variant (array of Double) (read/write)


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  24.1