The number of points is the number of measurements gathered in a sweep cycle in the range of stimulus change.

The number of points should be set for the linear and logarithmic frequency sweeps.

Increase the number of points to get a larger trace resolution. To increase measurement performance, reduce the number of points to values that provide an acceptable trace resolution. To maintain high accuracy, the number of points in the calibration and in the actual measurements must be the same.

The channel to which the function is applied must be preselected as active (See Selection of Active Channel).


To enter the number of points use the softkey Stimulus in the right menu bar. Then click on Points field, select the required value from the list and complete the setting by clicking Ok softkey.

Stimulus Point-2 Points



note

The number of Points can be set using the mouse (See Number of Points Setting).