SCPI Command
SENSe<Ch>:CORRection:COLLect[:ACQuire]:TRLLine <port1>,<port2>
Measures the calibration data of the TRL line standard between <port1> and <port2>. When the subclass points to the line standard (See SENS:CORR:COLL:SUBC) the command scans in both directions. When the subclass points to the match standard (TRM) the command scans in one direction and two commands with swapped <port1> and <port2> are required to complete the measurement. Note: The command starts the measurement immediately if the trigger source for calibration set to the "Internal" by the command SENS:CORR:TRIG:FREE, otherwise waits for the trigger signal. The command blocks the execution of the subsequent commands until the completion of the measurement. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...16}
<rcvport> |
The number of the receiver port from 1 to 2 |
<srcport> |
The number of the source port from 1 to 2 |
Out of Range
If the same port numbers are specified, an error occurs.