Please enable JavaScript to view this site.

 Network Analyzers using PXI software

SCPI Command

SENSe<Ch>:CORRection:COEFficient:SAVE

Description

Enables the written calibration coefficients depending on the selected calibration type. On completion of the command, the error correction automatically turns ON.

Executing this command before all necessary calibration coefficients have been written will result in an error and the command will be ignored.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Related Commands

Calibration type selection:

SENS:CORR:COEF:METH:ERES

SENS:CORR:COEF:METH:OPEN

SENS:CORR:COEF:METH:SHOR

SENS:CORR:COEF:METH:THRU

SENS:CORR:COEF:METH:SOLT1

SENS:CORR:COEF:METH:SOLT2

Calibration coefficient writing:

SENS:CORR:COEF

Equivalent Softkeys

None

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.3