SCPI Command

SENSe<Ch>:CORRection:COEFficient:METHod[:RESPonse]:THRU <rcvport>, <srcport>

Description

Selects the ports and sets the response calibration (Thru) type when the written calibration coefficients are made effective by the SENS:CORR:COEF:SAVE command.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<rcvport>

The number of the receiver port from 1 to 2

<srcport>

The number of the source port from 1 to 2

Out of Range

If the same port numbers are specified, an error occurs.

Related Commands

SENS:CORR:COEF:SAVE

Equivalent Softkeys